Rotary flow regulators have a standardized flow stability of +/- 10% over time (ISO8536-14 standard), and are therefore considered lt;precision lt; regulators. It should be noted that many factors influence flow regulation (viscosity, height, blood pressure, etc.): drop-counting and regular monitoring are therefore good practice when using these precision regulators. Purging information: graduations are reliable to +/- 20% from 20ml upwards, provided the following purging protocol is followed...: Purge the line while keeping the flow regulator in the lt;open lt; position (delivered in the lt;open lt; position). After purging the line, and before connecting the device to the patient's vascular access, turn the flow regulator to the lt;off lt; position. Contact us to obtain the correct use sheet.
Extender with :
High-quality tubing: flexible, anti-plication PVC
Precision rotary flow regulator: Smooth, easy adjustment - Constant, reliable flow thanks to graduated rotary flow regulator.
Optimum technical features: Movable Luer Lock tip
Pressure resistance: 4000 hPa (4 Bars - 58 psi)
Dead volume: 4.38ml.
Graduation from 0 to 250 mL/h.
Contact with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation time, while the luer lock is connected, and therefore under physical stress, presents a significant risk of luer lock cracking, and should be avoided wherever possible.
In line with SF2H (2019) recommendations, disinfecting connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective at disinfecting inert materials, and with much less risk of altering plastics.
Indications
This medical device is designed for the parenteral administration of injectable preparations (drugs, nutrients).
This device enables precise gravity administration of injectable parenteral preparations: drugs, solutions, parenteral nutrition.
Do not use with high-risk drugs or drugs with narrow therapeutic margins (catecholamines, morphine, heparin, insulin, potassium chloride, anticancer drugs requiring infusion pump administration, and all pharmaceutical specialties whose R.C.P. includes the need for infusion pump or syringe pump administration).
Source : OMEDIT CENTRE -- Les bonnes pratiques de perfusion : Module lt; La précision du débit de perfusion lt;.
Matériaux utilisés
Capuchon : PE
Luer Lock femelle : PC
Tube : PVC
Régulateur de débit : ABS / TPE
Site d'injection : ABS / Polyisoprène synthétique
Luer Lock mâle : ABS
Capuchon : PE
Étiquetage
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
dispositif médical
stérile
conformité
Conditions de transport
Protected from light and humidity
Conditionnement | Quantités de pièces | Dimensions en cm | Poids en kg | Ean 13 du conditionnement |
UNIT | 1 | 21 x 13 x 0 | 06972620110510 | |
BAG | 25 | 55 x 30 x 5 | 0.510 | |
CASE | 150 | 39 x 42 x 37 | 16972620110517 |