Rotary flow regulators have a standardized flow rate stability of +/- 10% over time (ISO8536-14 standard), and are therefore considered lt;precision lt; regulators. It should be noted that many factors influence flow regulation (viscosity, height, blood pressure, etc.)?Droplet counting and regular monitoring are therefore still good practice when using these precision regulators. This is why this rotary flow regulator is offered without graduation, to enable good compliance with these good practices. To ensure optimum adjustment, please follow the purging protocol below...: Purge the line with the flow regulator in the lt;open lt; position (delivered in the lt;open lt; position). After purging the line, and before connecting the device to the patient's vascular access, turn the flow regulator to the lt;off lt; position. Contact us to obtain the correct use sheet.
Prolonger with :
High-quality tubing: flexible, anti-plication PVC
Precision rotary flow regulator: Smooth, easy adjustment - Constant, reliable flow
Optimum technical features: Fixed and mobile Luer Lock tips
Pressure resistance: 4000 hPa (4 Bars - 58 psi)
Non-graduated flow regulator: No need to count drops
Made in China.
Contact with Chlorhexidine 0.5% or 2% in Isopropyl alcohol
Isopropyl alcohol at 70°, with no evaporation time, and while the luer lock
under physical stress, presents a significant risk of luer lock
of luer lock cracking, and should be avoided wherever possible.
In line with SF2H (2019) recommendations, disinfecting connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective at disinfecting inert materials and much less likely to damage plastics.
Indications
This device enables precise gravity administration of injectable parenteral preparations: drugs, solutions, parenteral nutrition.
Do not use with high-risk drugs, drugs with narrow therapeutic margins (catecholamines, morphine, heparin, insulin, potassium chloride, anticancer drugs requiring infusion pump administration, and all pharmaceutical specialties whose R.C.P. includes the requirement for administration by infusion pump or syringe pump, or via an infusion set fitted with a 0.2µm filter).
Source : OMEDIT CENTRE -- Les bonnes pratiques de perfusion : Module lt; La précision du débit de perfusion lt;.
To disinfect connections, use 70° alcohol, which is highly compatible. Increasingly, caregivers are using chlorhexidine 2% in 70% isopropyl alcohol. This type of solution is known to be very aggressive to plastics.
For this reason, Didactic conducted laboratory tests on the contact of connections with this type of disinfectant.
The results showed that the connections were fairly resistant to this composition when in relatively short contact (less than 1min) with a swab soaked in disinfectant solution without excess (1 to 2ml per swab).
Dispose of in accordance with local regulations.
Matériaux utilisés
Cap: PE
Female Luer Lock: PC
Tube: PVC
Flow regulator: ABS / TPE
Luer Lock male: ABS
Cap: PE
Étiquetage
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
dispositif médical
stérile
conformité
Conditions de transport
Protected from light and humidity
Conditionnement | Quantités de pièces | Dimensions en cm | Poids en kg | Ean 13 du conditionnement |
UNIT | 1 | 21 x 13 x 0 | 06972620110527 | |
BAG | 25 | |||
CASE | 150 | 39 x 42 x 31 | 16972620110524 |