You are:
The DEHP-free, BPA-free INFINEED lt;Gravity lt; 1-way infusion set was designed to minimize the device's environmental impact. Two methods have been used: first, by reducing the plastic mass used by around 10%, without making any changes to functionality; second, by exclusively using raw materials that are free from endocrine disruptors (elimination of components comprising DEHP or BPA.) Residual volume: 15ml
For disinfection of connections, preferably use 70° alcohol which ensures very good compatibility. More and more often, caregivers are using 2% chlorhexidine in 70% isopropyl alcohol. This type of solution is known to be very aggressive for plastic materials.
For this reason Didactic has conducted laboratory tests of contact connections with this type of disinfectant agent.
The results showed that the connections provided fairly good resistance to this composition when the contact is relatively short (less than 1 minute) with a compress soaked in disinfectant solution without excess (1 to 2ml per compress).
Two-channel perforator
Air intake: Delivered closed for immediate use on a 0.2µm filter bag, Chamber: Chamber: 5.5 mL / 5 cm long - Flexible and transparent, Calibrated at 20 drops = 1ml ± 0.1ml. With 15µm particle filter.
Flow regulator: ABS linear flow control for good roller clamp rigidity, precision and reliability.
Tubing: PVC (DINCH) designed for comfort, optimal flow adjustment consistency and to prevent kinks.
End fitting: Mobile Luer Lock, end stopper equipped with a purge filter with anti-bacterial membrane. Very pressure-resistant purge filter.
Length: 160cm
Indications
This device enables the parenteral administration by gravity of injectable preparations: parenteral drugs, solutions and nutrition.
For peripheral lines, the infusion line can remain in place for a maximum of 96 hours (i.e. the maximum time the peripheral catheter can be in place).
For central lines, the infusion line must be changed at least every 72 hours (or according to the institution's validated protocol).
Gravity infusion set, do not use for pressure infusion, transfusion or medicine that need a 0.2µm filter.
Dispose according to the state member regulation.
Used materials
PVC (DEHP < 01.% (m/m)): Chamber, air intake, tubing - DINCH: plasticizer - HDPE: protective perforator cap, end cap - Acrylonitrile Butadiene Styrene: Perforator, linear flow regulator, roller, mobile Luer Lock - Polyamide: Particle filter - Polyethylene/PTFE: Purge filter
Labeling
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
medical device
sterile
conformity
economic operator
Transport conditions
Protect from light and humidity and store at a temperature between 5°C and 20°C.
Conditioning | Parts quantities | Dimensions in cm | Weight in kg | Ean 13 of conditioning |
UNIT | 1 | 10 x 21 x 1 | 0.029 | 3661809028939 |
BAG | 25 | 48 x 27 x 4 | 0.660 | 3661809128936 |
CASE | 250 | 59 x 39 x 27 | 7.624 | 3661809228933 |
PALLET | 6000 |